Power problem / laser source failure

uitval-laserbron

After checking the software of the laser source, the water temperature at full power appeared to rise rapidly. After 3 minutes a temperature alarm in the source was activated. However, there were no other indications that indicated the cause of this increase. The flow seemed to be fine. No excessive pump currents/voltages or other electronic causes could be found.

By physically checking the cooling lines we noticed that one of the modules was clearly getting warm while the others were staying cool. The suspicion was that there was an obstruction in this module which caused the flow in this single module to be reduced.

A few technical inputs later this suspicion turned out to be correct. It concerned a piece of tape, which probably had not been removed sufficiently from the pipe during installation, or had gotten into the pipe during transport. That this phenomenon only occurred after a reasonable number of operating hours is probably because this piece of tape slowly moved further into the system until it finally had a real effect on the operation.

After the intervention the temperatures remained at normal temperature values ​​and remained nicely at temperature at full power. Problem solved…